Service technicians may:
- (a) Remove, in conjunction with repairing, rebuilding, or reconditioning devices, official security seals or rejection tags or marks placed on the devices by the division;
- (b) Place devices into commercial service if they meet the requirements of NIST Handbook 44, unless ordered not to do so by the division;
(c) Install devices:
- (1) In accordance with NIST Handbook 44, the instructions of the device manufacturer and any instructions marked on the devices; and
- (2) In fixed locations so that their operation and performance will not be adversely affected by any characteristics of the foundation, supports or other aspect of their installation;
- (d) Adjust devices by making all adjustments and calibrations to devices as closely as practicable to zero error; and
- (e) Remove and appropriately affix security seals to any adjustment mechanisms designed to be sealed, unless ordered otherwise by the division.
Source. (See Revision Notes #1 and #2 at chapter heading for Agr 1400) #12424, eff 12-1-17