PURPOSE: This rule establishes special requirements for mobile medical radiographic installations.
- (1) All mobile equipment shall be provided with cones or frames so that the minimum target-to-skin distance is at least eighteen inches (18").
- (2) The exposure control switch shall be so arranged that the operator can stand at least six feet (6') from the patient and outside the primary beam for all exposures.
- (3) A device shall be provided to terminate the exposure after a preset time.
AUTHORITY: section 192.420, RSMo 1986.* This rule was previously filed as 13 CSR 50- 90.160. Original rule filed Nov. 9, 1964, effective Dec. 9, 1964. *Original authority: 192.420, RSMo 1963.