310 C.M.R. 42.08
(1) Personnel. Certified laboratories and laboratories seeking certification shall designate a laboratory director and laboratory supervisor(s), and employ a sufficient number of qualified analysts commensurate with the laboratory's workload. The designated laboratory director and laboratory supervisor may be the same individual if he or she possesses the minimum qualifications and fulfills the responsibilities of both positions as set in 310 CMR 42.08(1).
(a) The laboratory director shall have the following responsibilities:
(b) A laboratory supervisor shall have the following responsibilities:
(c) Minimum Qualifications of Laboratory Director.
(d) Minimum Qualifications of Laboratory Supervisor for Laboratories Certified in the Discipline of Chemistry or Radiochemistry.
2. Experience.
(e) Minimum Qualifications of Laboratory Supervisor for Laboratories Certified in the Discipline of Microbiology.
(f) Minimum Qualifications of Instrumentation Analyst.
(g) Minimum Qualifications of Analyst.
(2) Facilities. A certified laboratory shall have:
(3) Equipment and Materials. Certified laboratories and laboratories seeking certification shall have readily available on the premises all equipment, supplies, reagents, glassware, and instrumentation necessary to perform the analyses for which the laboratory is certified or seeks certification and related quality control activities. Such equipment and materials shall be maintained in good working condition, meet the performance criteria of the analytical method used, and, where appropriate, meet the criteria specified in 310 CMR 42.08(3):
(5) Quality Assurance/Quality Control.
(a) General Requirements.
1. Quality Assurance Plan. Certified laboratories shall establish, maintain, and follow a written Quality Assurance (QA) Plan acceptable to the Department. Each laboratory's QA plan shall be made available to all analysts employed by the laboratory. At a minimum, QA plans shall include:
c. Standard operating procedures (SOPs) that accurately reflect all phases of current laboratory activities. The standard operating procedures section of a laboratory's QA plan shall include the following:
procedure may be a separate document. The QA plan must include a list of all such standard operating procedures. Only the laboratory director or supervisor may make changes in standard operating procedures. Such changes shall be effective only when put in writing. d. Record Maintenance. The record maintenance procedures section of a laboratory's QA plan shall include the procedures for creating, controlling, and maintaining the following records:
2. Temperature Records.
3. Laboratory Chemicals and Reagents. Analytical reagent (AR) grade or American Chemical Society (ACS) grade chemicals or better are required for analyses, unless otherwise required by the analytical method. In addition, laboratory chemicals and reagents shall meet the following requirements:
6. Instrumentation Calibration Requirements.
a. General Requirements.
c. pH Meters. Each pH meter shall be calibrated daily or prior to each use with pH 7.0 and pH 4.0 or 10.0 buffer standards that bracket the expected value of the sample, medium, or reagent being tested. The buffers used shall be recorded, including the date of calibration and the name of the analyst who performed the calibration.
d. Conductivity Meters.
e. Thermometers. The accuracy of all temperature measurement devices used to monitor temperatures shall be verified by comparing the reading of each device with that of a certified reference thermometer that is graduated in degree increments no larger than those of the device whose accuracy is being verified. The laboratory must discontinue use of a thermometer graduated in 0.5(cid:47)C increments or less that differs from the certified thermometer by more than 1(cid:47)C. The accuracy of glass and electronic thermometers must be verified annually; metal thermometers must be verified quarterly; infrared detection devices must be verified every six months; and the certified reference thermometer must be calibrated at least once every five years. The correction factor and date of verification of accuracy must be indicated on the thermometers. The laboratory shall maintain a secure record that includes:
7. Sample Collection, Preservation and Handling.
e. Chain-of-custody information must include:
(b) Additional Requirements for Chemical and Radiochemical Laboratories.
1. Quality Control Procedures.
e. When integrating chromatography peaks, either automatically or manually, each laboratory must ensure that integrations are performed in a correct and consistent manner for standards and samples, including quality control samples. Each laboratory must maintain documentation of manual integrations that includes the following:
2. Determination of Method Detection Limit.
(c) Additional Requirements for Microbiology Laboratories.
1. Autoclaves.
3. Incubators.
6. Sterility of Rinse/Dilution Water and Sample Bottles.
7. Residue Testing of Glassware.
8. Microbiological Media - Quality Control Measures.
10. Membrane Filter Procedure Quality Control Specifics.
11. Quality Control.
12. Laboratory Reagent Water. The laboratory shall use satisfactorily tested reagent water from a water purification system (e.g., still, deionization unit, or a reverse-osmosis unit) to prepare media, reagents, and dilution/rinse water for performing microbial analyses.
e. The test of the bacteriological quality of the laboratory reagent water must be performed annually. The ratio of the growth rate must be between 0.8 and 3.0. The
th th
test is described in Section 9020B of the 18 and 19 editions of Standard Methods for the Examination of Water and Wastewater. The bacteriological quality test is not required of laboratories that document that their laboratory reagent water meets the
th th
criteria for Type II water or better, as defined in Standard Methods (18 and 19 editions), Section 1080C or Medium quality water or better as defined in Standard
th
Methods (20 edition), Section 1080C.