Ill. Admin. Code tit. 32, § 340.310
a) Each licensee or registrant shall conduct operations so that:
2) The total effective dose equivalent to individual members of the public from a radiation machine does not exceed:
B) 1 mSv (0.1 rem) in any year at any location within a facility where a radiation machine is installed or where the radiation machine or its use changes on or after January 1, 1994.
AGENCY NOTE: It is the Agency's intent to allow registrants using radiation machines in facilities designed to the 5 mSv (0.5 rem) limit to continue to use the 5 mSv (0.5 rem) total effective dose equivalent limit for a member of the public. This includes locations where the intensity of the radiation machine is not increased beyond the design basis, the type of radiation machine use is not changed and the type of facility use is not changed; or
B) Any medical administration the individual has received;
D) Voluntary participation in medical research programs; and
b) A licensee may apply for prior Agency authorization to operate up to an annual dose limit for an individual member of the public of 5 mSv (0.5 rem). This application shall include the following information:
3) The procedures to be followed to maintain the dose ALARA.
(Source: Amended at 29 Ill. Reg. 20841, effective December 16, 2005)