Ill. Admin. Code tit. 32, § 333.40
a) The Department provides the following calibration services under accreditation by the Conference of Radiation Control Program Directors, Inc.
1) X-ray Calibrations. 5 beam codes at 32 mR/sec. This service is available for the Radcal MDH 1015 x-ray monitor only. The available beam codes are:
| BEAM CODE | HALF-VALUE LAYER | HOMOGENEITY COEFFICIENT | |
| L | 80 | 1.81 mm Aluminum | 0.59 |
| L | 100 | 2.82 mm Aluminum | 0.59 |
| M | 30 | 0.36 mm Aluminum | 0.65 |
| M | 50 | 1.03 mm Aluminum | 0.64 |
| M | 100 | 4.96 mm Aluminum | 0.73 |
2) Gamma Ray Calibrations. 1 beam code. This service is available only for ion chamber and energy compensated G-M tube instruments.
| BEAM CODE | INTENSITIES |
| Cs-137 | .005 - 40 R/hr |
b) The Department also provides the following services, which are not performed under accreditation.