For blood gas analyses, the laboratory must perform the following:
- (a) Calibrate or verify calibration according to the manufacturer's specifications and with at least the frequency recommended by the manufacturer.
- (b) Test one sample of control material each 8 hours of testing using a combination of control materials that include both low and high values on each day of testing.
- (c) Test one sample of control material each time specimens are tested unless automated instrumentation internally verifies calibration at least every 30 minutes.
- (d) Document all control procedures performed, as specified in this section.