42 C.F.R. § 84.103
(c) When the ambient temperature during testing is 24 ±6 °C. (75 ±10 °F.), the maximum temperature of inspired air recorded during man tests shall not exceed the following, after correction for deviation from 24 °C. (75 °F.):
| Where service life of apparatus is— | Where percent relative humidity of inspired air is— | Maximum permissible temperature of inspired air shall not exceed— | |
|---|---|---|---|
| °F. | °C. | ||
| 1⁄4 hour or less | 0-100 | 135 | 57 |
| 1⁄4 hour to 3⁄4 hour | 0-5050-100 | 1251 110 | 521 43 |
| 1 to 2 hours | 0-5050-100 | 1151 105 | 461 41 |
| 3 hours | 0-5050-100 | 1101 100 | 431 38 |
| 4 hours | 0-5050-100 | 1051 95 | 411 35 |
| 1 Where percent relative humidity is 50-100 and apparatus is designed for escape only, these maximum permissible temperatures will be increased by 5 °C (10 °F). |