- (a) Identification. An esophageal tissue characterization system is a device intended for obtaining measurements of electrical properties within esophageal tissue.
(b) Classification. Class II (special controls). The special controls for this device are:
- (1) All patient contacting components of the device must be demonstrated to be biocompatible.
- (2) Performance testing must demonstrate the device can accurately measure the designated electrical characteristics.
- (3) Mechanical safety testing must demonstrate that the device will withstand forces encountered during use.
- (4) Software verification, validation, and hazard analysis must be performed.
- (5) Electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety of the device must be performed.
- (6) Performance data must validate the reprocessing instructions for any reusable components of the device.
(7) Labeling must include:
- (i) Specific instructions regarding the proper placement and use of the device;
- (ii) Instructions for reprocessing of any reusable components; and
- (iii) An expiration date for single use components.
[86 FR 68399, Dec. 2, 2021]