21 C.F.R. § 606.60
(b) Equipment that shall be observed, standardized and calibrated with at least the following frequency, include but are not limited to:
| Equipment | Performance check | Frequency | Frequency of calibration |
|---|---|---|---|
| Temperature recorder | Compare against thermometer | Daily | As necessary. |
| Refrigerated centrifuge | Observe speed and temperature | Each day of use | Do. |
| Hematocrit centrifuge | Standardize before initial use, after repairs or adjustments, and annually. Timer every 3 mo. | ||
| General lab centrifuge | Tachometer every 6 mo. | ||
| Automated blood-typing machine | Observe controls for correct results | Each day of use | |
| Hemoglobinometer | Standardize against cyanmethemoglobin standard | ......do | |
| Refractometer | Standardize against distilled water | ......do | |
| Blood container scale | Standardize against container of known weight | ......do | As necessary. |
| Water bath | Observe temperature | ......do | Do. |
| Rh view box | ......do | ......do | Do. |
| Autoclave | ......do | Each time of use | Do. |
| Serologic rotators | Observe controls for correct results | Each day of use | Speed as necessary. |
| Laboratory thermometers | Before initial use. | ||
| Electronic thermometers | Monthly. | ||
| Vacuum blood agitator | Observe weight of the first container of blood filled for correct results | Each day of use | Standardize with container of known mass or volume before initial use, and after repairs or adjustments. |
[40 FR 53532, Nov. 18, 1975; 40 FR 55849, Dec. 2, 1975, as amended at 45 FR 9261, Feb. 12, 1980; 57 FR 11263, Apr. 2, 1992; 57 FR 12862, Apr. 13, 1992]