Cal. Code Regs. tit. 17, § 30309
Special Requirements for Mobile Radiographic Equipment.
Effective Mar 15, 2018Register 2018, No. 11Authority cited: Sections 114975, 115000, 115060 and 131200, Health and Safety Code. Reference: Sections 114965, 114970, 115060, 131050, 131051 and 131052, Health and Safety Code.State of California
(a) Equipment.
- (1) All requirements of Section 30308(a) apply except 30308 (a)(5) and 30308 (a)(9).
- (2) The exposure control switch shall be of the dead-man type and shall be so arranged that the operator can stand at least 6 feet from the patient and well away from the useful beam.
- (3) Inherent provisions shall be made so that the equipment is not operated at source-skin distances of less than 12 inches.
(b) Operating Procedures.
- (1) All provisions of Section 30308(b) apply except 30308(b)(5).
- (2) The target-to-skin distance shall be not less than 12 inches.
- (3) Personnel monitoring shall be required for all individuals operating mobile X-ray equipment.
Note: Authority cited: Sections 114975, 115000, 115060 and 131200, Health and Safety Code. Reference: Sections 114965, 114970, 115060, 131050, 131051 and 131052, Health and Safety Code.
History
1. Amendment filed 3-5-71; effective thirtieth day thereafter. Approved by State Building Standards Commission 2-26-71. (Register 71, No. 10).
2. Renumbering and amendment filed 9-4-73 as an emergency; effective upon filing (Register 73, No. 36). Approved by State Building Standards Commission 11-30-73.
3. Certificate of Compliance filed 12-28-73 (Register 73, No. 52).
4. Amendment of subsection (b)(1) filed 6-18-87; operative 7-18-87 (Register 87, No. 28).
5. Amendment of Note filed 3-15-2018; operative 3-15-2018. Submitted to OAL for filing and printing only pursuant to Health and Safety Code section 100425 (Register 2018, No. 11).