A. A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform spot-checks of each gamma stereotactic radiosurgery facility and on each unit:
- 1. Monthly;
- 2. Before the first use of the unit on a given day; and
- 3. After each source installation.
B. A licensee shall:
- 1. Perform the measurements required by subsection (A) in accordance with written procedures established by an authorized medical physicist. That individual need not actually perform the spot-check measurements.
- 2. Have the authorized medical physicist review the results of each spot-check within 15 days. The authorized medical physicist shall notify the licensee as soon as possible in writing of the results of each spot-check.
C. To satisfy the requirements of subsection (A)(1), spot-checks shall, at a minimum:
1. Assure proper operation of:
- a. Treatment table retraction mechanism, using backup battery power or hydraulic backups with the unit off;
- b. Helmet microswitches;
- c. Emergency timing circuits; and
- d. Stereotactic frames and localizing devices (trunnions).
2. Determine:
- a. The output for one typical set of operating conditions measured with the dosimetry system described in R9-7-733(B);
- b. The difference between the measurement made in subsection (C)(2)(a) and the anticipated output, expressed as a percentage of the anticipated output;
- c. Source output against computer calculation;
- d. Timer accuracy and linearity over the range of use;
- e. On-off error; and
- f. Trunnion centricity.
D. To satisfy the requirements of subsections (A)(2) and (A)(3), spot-checks shall assure proper operation of:
- 1. Electrical interlocks at each gamma stereotactic radiosurgery room entrance;
- 2. Source exposure indicator lights on the gamma stereotactic radiosurgery unit, on the control console, and in the facility;
- 3. Viewing and intercom systems;
- 4. Timer termination;
- 5. Radiation monitors used to indicate room exposures; and
- 6. Emergency off buttons.
- E. A licensee shall arrange for the repair of any system identified in subsection (C) that is not operating properly as soon as possible.
- F. If the results of the checks required in subsection (D) indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
- G. A licensee shall retain a record of each check required by subsections (C) and (D) for three years from the date of the procedure, and a copy of the procedures required by subsection (B) until licensee terminates all medical activities involving the radiosurgery unit.
Historical Note
New Section R9-7-739 recodified from R12-1-739 at 24 A.A.R. 813, effective March 22, 2018 (Supp. 18-1).