OPINION
After their newborn daughter was diagnosed with a heart defect, was given a Guidant pacemaker, and tragically passed away at the age of fourteen months, the appellants filed a state law complaint against Guidant. Among other things, they argue that Guidant should have put specific warnings on the pacemaker labeling related to its implantation into pediatric patients. Because the label had been preapproved by the Food and Drug Administration, however, and Guidant was not required to include the additional warnings, the trial court held that any state law-based failure-to-warn claims were preempted by federal law. Finding that the trial court properly found the claims preempted, we affirm.
Appellants-plaintiffs Jodi McGookin, as mother of the deceased, Samantha Arlene McGookin, Vicky McGookin, Jim McGoo-kin, and Julian E. Smith (collectively, the
FACTS
Samantha’s Story
Samantha was born on April 27, 2004, with complete heart block, which is a disorder of the heart’s electrical system. Complete heart block occurs in people of all ages, and both children and adults with the condition are treated with pacemakers. Pacemakers are medical devices designed to cause the human heart to beat by providing low-voltage electrical impulses delivered from the device through wires, called “leads,” connected to the heart. On April 30, 2004, a Guidant Insignia I Ultra Model 1290 (Insignia 1290) was implanted in Samantha’s abdomen and attached to her heart with two unipolar and epicardial leads. Medical records throughout Samantha’s life, and testing conducted by Guidant after her death, show that her device provided therapy at all times. On July 9, 2005, Samantha died.
The Insignia 1290
The United States Food and Drug Administration (FDA) must approve all Class III medical devices — such as the Insignia 1290 — before they may be commercially distributed. On August 15, 2008, Guidant filed a supplemental application for several devices, including the Insignia 1290. The application contained detailed information regarding the Insignia 1290, including its design, testing, indications for use, contraindications, warnings and precautions, and its Automatic Capture 1 feature. The application also stated that, consistent with industry practice, pediatric patients had not been included in clinical trials. On November 4, 2008, the FDA approved the Insignia 1290 as a class III medical device, and that approval remains effective today. The FDA’s approval included approval of the Insignia 1290 labeling, which is the same today as it was in November 2003.
At the heart of the Appellants’ case is their claim that the labeling for Samantha’s pacemaker was inadequate because it failed to warn of a lack of testing of the Automatic Capture feature with small children, unipolar epicardial leads, and abdominal implantation. In other words, their complaint challenges labeling expressly approved by the FDA.
On August 25, 2006, the McGookins filed a complaint against Guidant for wrongful death, product liability, breach of express and implied warranties, actual fraud, constructive fraud, negligence, violation of the Indiana Deceptive Consumer Sales Act, and intentional and negligent infliction of emotional distress. Julian Smith, Samantha’s father, filed similar claims against Guidant, adding a claim for negligence per se.
On April 1, 2009, Guidant moved for summary judgment based on federal preemption, lack of evidence of defect or causation, and certain claim-specific reasons. On July 22, 2009, the trial court issued an
To the extent plaintiffs seek to impose liability on the defendants under substantive legal theories on the basis that the Insignia 1290 ... should have been subject to requirements and specifications in addition to or different from the device specific requirements imposed by federal law, including specifically the [FDA] regulations, plaintiffs’ claims are preempted by the Medical Device Amendments of 1976 (“MDA”). To the extent plaintiffs seek to enforce or restrain violations of federal law including the MDA, their claims are preempted. To the extent plaintiffs seek to impose liability on the defendants under substantive legal theories that seek monetary damages predicated on challenges to conduct of the defendants allowed by and not in violation of any applicable federal requirements including FDA regulations plaintiffs’ claims are preempted.
Appellants’ App. p. 22. To the extent that the Appellants’ claims were premised on a violation of FDA regulations, the trial court held that the claims were not preempted. The trial court went on to find many genuine issues of material fact as to each of the Appellants’ claims except for the implied warranty claim, and it merged the claims based on negligence, negligence per se, actual and constructive fraud, intentional and negligent infliction of emotional distress, and the Indiana Deceptive Consumer Sales Act, into a single cause of action under the Indiana Products Liability Act. The trial court then entered the following order:
1. Defendants’ Motion for Summary Judgment is granted as to Plain-tiffsf] claims based on breach of implied warranty. ...
2. Defendants^] Motion for Summary Judgment is granted as to any claims of plaintiffs that are not premised upon a violation of an applicable federal requirement except for claims based upon an assumed duty and breach of express warranty resulting from representations made by defendants to plaintiffs or plaintiffs’ physicians that were not true.
3. In all other respects Defendants’ Motion for Summary Judgment is denied.
Id. at 27-28. In a subsequent order, the trial court clarified that the claims folded into the single Indiana Products Liability Act cause of action had not been dismissed.
A ten-day jury trial took place between August 25 and September 4, 2009. All of the Appellants’ claims aside from implied warranty and manufacturing defect 2 were litigated during the trial. On September 4, 2009, the case was submitted to the jury, with instructions on six substantive claims: product liability based on failure to warn, negligence (based on failure to warn and design defect), negligent and intentional infliction of emotional distress, breach of assumed duty, and breach of express warranty. 3 The jury found in favor of Guidant on all claims on September 8, 2009.
On October 8, 2009, the Appellants filed a motion to correct error, alleging, among other things, that the trial court had erred
DISCUSSION AND DECISION
I. Standard of Review
The Appellants are appealing the denial of their motion to correct error, which alleged that the trial court erred in granting a portion of Guidant’s motion for summary judgment. We review a trial court’s ruling on a motion to correct error for an abuse of discretion.
Newland Resources, LLC v. Branham Corp.,
Summary judgment is appropriate only if the pleadings and evidence considered by the trial court show that there is no genuine issue as to any material fact and that the moving party is entitled to judgment as a matter of law.
Owens Corning Fiberglass Corp. v. Cobb,
II. Preemption
The Appellants contend, based on a series of federal and Indiana cases that will be explored below, that the trial court erroneously held that the Appellants’ claims were preempted unless the claims were premised on a violation of federal statutes and/or regulations. The substantive crux of their argument is as follows:
... Although Guidant’s label complied with the FDA requirements of its pre-market approval, other FDA regulations gave Guidant the ability to add to or strengthen those regulations without prior FDA approval. The Indiana Product Liability Act incorporates a “reasonableness” component in determining whether warnings are inadequate. Therefore, it becomes a jury question as to whether Guidant acted reasonably in failing to add to or strengthen its warnings pursuant to 21 C.F.R. § 814.39(d). The label, for example, could have informed consumers and physicians that the pacemaker had not been tested in infants or with epicardial leads, or with an abdominal implant. The label could have stated that use with epicardial leads in infants was contraindicated. The trial court therefore erred in granting Guidant’s motion for summary judgment and ruling that any attempt to impose liability on Guidant under substantive legal theories ... predicated on challenges to conduct of Guidant allowed by and not in violation of any applicable federal requirements are preempted.
Appellants’ Br. p. 26.
A. Riegel
Before this case went to trial, the United States Supreme Court decided
Riegel v. Medtronic, Inc.,
Except as provided in subsection (b)[ 5 ] of this section, no State or political subdivision of a State may establish or continue in effect with respect to a device intended for human use any requirement—
(1) which is different from, or in addition to, any requirement apphcable under this chapter to the device, and
(2) which relates to the safety or effectiveness of the device or to any other matter included in a requirement apphcable to the device under this chapter.
21 U.S.C. § 360k(a).
At issue in
Riegel
was a balloon catheter, a Class III medical device that received premarket approval from the FDA. After the catheter was inserted in Charles Riegel’s coronary artery, serious complications ensued, and Riegel and his wife filed a lawsuit alleging that the catheter was designed, labeled, and manufactured in a manner violating New York common law. To determine whether the Riegels’ claims were preempted by the MDA, the Court first considered whether the federal government had established requirements ap-phcable to the catheter. Answering in the affirmative, the Court held that premarket approval, synonymous with federal safety review, does impose “requirements” under the MDA.
Id.
at 322-23,
Next, the Riegel Court turned to a second question:
whether the Riegels’ common-law claims rely upon “any requirement” of New York law apphcable to the catheter that is “different from, or in addition to” federal requirements and that “relates to the safety or effectiveness of the device or to any other matter included in a requirement apphcable to the device.”
Id.
at 323,
B. Wyeth
In 2009 — before the parties briefed the preemption issue on summary judgment herein — the United States Supreme Court decided
Wyeth v. Levine, 555
U.S. 555, 129 5.Ct. 1187,
The
Wyeth
Court held that federal law did not preempt the patient’s failure-to-warn claim. Relying heavily on the nature of the Food, Drug, and Cosmetic Act
6
(FDCA), the Court explicitly noted that unlike the MDA, the FDCA does not contain an express preemption provision. 129
The drug manufacturer contended “that the FDCA establishes both a floor and a ceiling for drug regulation: Once the FDA has approved a drug’s label, a state-law verdict may not deem the label inadequate, regardless of whether there is any evidence that the FDA has considered the stronger warning at issue.” Id. at 1199. The Court rejected this argument, noting that
Congress enacted the FDCA to bolster consumer protection against harmful products_If Congress thought state-law suits posed an obstacle to its objectives, it surely would have enacted an express pre-emption provision at some point during the FDCA’s 70-year history.... Its silence on the issue, coupled with its certain awareness of the prevalence of state tort litigation, is powerful evidence that Congress did not intend FDA oversight to be the exclusive means of ensuring drug safety and effectiveness.
Id. at 1199-1200. Having examined the nature of the FDCA and the FDA’s traditional position on these issues, the Court concluded that “Wyeth has not persuaded us that failure-to-warn claims like Levine’s obstruct the federal regulation of drug labeling,” and held that Levine’s state law claims were not preempted by the FDCA. Id. at 1204.
C. Cook
Following
Wyeth,
the trial court herein issued its summary judgment order and the jury trial was held. After the conclusion of the trial-, but before the motion to correct error was due, a panel of this court issued
Cook v. Ford Motor Co.,
When a motor vehicle safety standard is in effect under this chapter, a State or a political subdivision of the State may prescribe or continue in effect a standard applicable to the same aspect of performance of a motor vehicle or motor vehicle equipment only if the standard is identical to the standard prescribed under this chapter.
49 U.S.C. § 30103(b)(1).
Additionally, however, unlike the MDA, the Safety Act contains a state common law savings clause: “[cjompliance with a motor vehicle safety standard under this chapter does not exempt a person from liability at common law.” 49 U.S.C. § 30103(e). In other words, “common law tort actions are not expressly preempted.”
Cook,
D. Appellants’ Claims
The Appellants argue that
Cook
establishes that the
Wyeth
rule can be applied to regulatory schemes aside from just the FDCA. While that is true, we cannot agree that
Cook
necessarily means that
Wyeth
applies to the MDA and class III medical devices. As noted above, the
Wyeth
Court explicitly distinguished the FDCA from the MDA, inasmuch as the FDCA has no preemption clause. And as for
Cook,
we note that while the Safety Act has a preemption clause, it also has an explicit savings clause that “preserves those actions that seek to establish a greater safety than the minimum safety achieved by a federal regulation intended to provide a floor.’ ”
Cook,
The MDA and
Riegel
could not be clearer that federal law broadly preempts any claim that would allow a jury to impose a standard of care different from or in addition to the FDA’s specific federal requirements.
Riegel,
The judgment of the trial court is affirmed.
Notes
. The Automatic Capture feature is designed to save battery life by measuring a patient’s capture threshold — the minimum amount of voltage required to cause the heart to beat— and then adjusting the output voltage to a set level above that threshold. The label, as approved by the FDA, states that this feature "is designed to work with any ventricular lead.” Appellees’ App. p. 1117.
. The Appellants abandoned the manufacturing defect claim before trial.
. Before the case was submitted to the jury, the Appellants voluntarily withdrew their claims for actual and constructive fraud, violation of the Indiana Deceptive Consumer Sales Act, and negligence per se.
. 21 U.S.C. § 360c et seq.
. The exception contained within subsection (b) permits the FDA to exempt certain state
. 21 U.S.C. § 301 etseq.
. 49 U.S.C. § 30101 etseq.
. Because we have ruled in Guidant’s favor, we need not address its contingent cross-appeal.