ORDER
This cause comes before the Court on Defendant Medtronic, Inc.’s (“Defendant”) Dispositive Motion to Dismiss Plaintiff Peggy McClelland’s (“Plaintiff’) Second Amended Complaint (“Motion to Dismiss”) (Doc. 45). Plaintiff filed a response in opposition to Defendant’s Motion to Dismiss (“Response”) (Doc. 46), to which Defendant replied (Doc. 52). For the following reasons, Defendant’s Motion to Dismiss will be granted.
I. BACKGROUND
A. Statement of Facts
Breanne A. McClelland (the “Decedent”) was a resident of Orange County, Florida until her death on August 11, 2011. Doc. 44, ¶ 1. Plaintiff is the Decedent’s mother and the duly appointed personal representative of the Decedent’s estate. Id. at ¶ 2. Defendant is a corporation organized under the laws of Minnesota, with its principal place of business in that State. Id. at ¶ 4; Doc. 1, ¶ 12. Defendant “researches, develops, tests, manufactures, markets, promotes, advertises, and sells medical devices.” Doc. 44, ¶ 6. Defendant’s largest business segment, Cardiac Rhythm Disease Management, develops products that restore and regulate heart rhythm and improve the heart’s pumping function. Id.
In March 2004, Defendant received premarket approval (“PMA”) from the U.S. Food and Drug Administration (“FDA”) for the EnPulse pacemaker Model E1DR21 (the “E1DR21”). Id. at ¶8. On or about April 30, 2004, the E1DR21 was surgically implanted in the Decedent. Id. at ¶ 9. Plaintiff alleges that prior to 2009, Defendant became aware that several of its implantable pulse generators, including the E1DR21, were defective in that, they could “lock-up and inappropriately trigger ERI/RRT mode in the pacemakers,” leading to “intense cardiac symptoms and failfure] to properly regulate cardiac rhythm.” Id. at ¶ 10.
B. Procedural History
On August 4, 2011, Plaintiff filed a Complaint in the Circuit Court of the Ninth Judicial Circuit in and for Orange County, Florida, asserting state law claims for negligence and strict liability against Defendant. See Doc. 2. Defendant subsequently removed the case to this Court based on diversity jurisdiction under 28 U.S.C. § 1332. See Doc. 1. Thereafter, Plaintiff filed an Amended Complaint against Defendant asserting state law claims for negligence per se and failure to warn. See Doe. 20. Defendant filed a motion to dismiss the Amended Complaint, to which Plaintiff responded. See Docs. 24, 26. On September 27, 2012, this Court entered an Order granting Defendant’s motion to dismiss, holding that Plaintiff had failed to state a claim upon which relief could be granted, and that Plaintiffs claims were barred by express and implied preemption. See Doc. 43,
On October 11, 2012, Plaintiff filed her Second Amended Complaint, asserting a single claim for state law negligence. See Doc. 44, p. 7. Plaintiff alleges that Defendant had a duty to notify the FDA of defects in the E1DR21 which could cause or contribute to death or serious injury, and that Plaintiff breached this duty and violated federal laws and regulations by “failing to promptly and accurately report to the FDA incidents and problems with the EnPulse E1DR21 that occurred after the device received [PMA].” Id. at ¶¶ 28-32. Plaintiff further alleges that if Defendant had complied with its duty to notify the FDA, the E1DR21 would have been recalled prior to the death of the Decedent. Id. at ¶ 34. The instant Motion to Dismiss followed.
II. STANDARD OF REVIEW
To survive a motion to dismiss, a pleading must include a “short and plain statement of the claim showing that the pleader is entitled to relief.” Ashcroft v. Iqbal,
III. DISCUSSION
A. Statutory Framework of the Medical Device Amendments
The regulation of medical devices entering the market is governed by the Federal Food, Drug, and Cosmetic Act (“FDCA”), 21 U.S.C. § 301 et seq. The FDCA provides that the enforcement of violations “shall be by and in the name of the United States.” 21 U.S.C. § 337(a). In 1976, Congress passed the Medical Device Amendments of 1976 (“MDA”), which amended the FDCA and imposed a regime of detailed federal oversight. See 21 U.S.C. § 360c et seq. The new regulatory regime separated medical devices into three classes based on the risks they pose to the public. Id.; Riegel v. Medtronic, Inc.,
1. Premarket Approval
All Class III devices, such as the E1DR21 manufactured by Defendant, must undergo the “rigorous” PMA process administered by the FDA. Medtronic, Inc. v. Lohr,
Even after PMA is granted, manufacturers are forbidden to make changes in design specifications, manufacturing processes, labeling, or any other attribute that would affect safety or effectiveness, without FDA permission. Id. (citing 21 U.S.C. § 360e(d)(6)(A)(i)). If a manufacturer wishes to make such a change to a device, it must submit an application for supplemental PMA, which is evaluated under
2. Preemptive Effect of the MDA
Until the statutory enactment of the MDA, the introduction of new medical devices was left largely for the states to supervise as they saw fit. Id. at 315,
[N]o State or political subdivision of a State may establish or continue in effect with respect to a device intended for human use any requirement—
(1) which is different from, or in addition to, any requirement applicable under this chapter to the device, and
(2) which relates to the safety or effectiveness of the device or to any other matter included in a requirement applicable to the device under this chapter.
21 U.S.C. § 360k(a). In Riegel, the Supreme Court suggested that the MDA’s preemption of certain state obligations in favor of more detailed federal oversight is justified because of the harm that would be caused by stifling innovation in medical devices if “juries were allowed to apply the tort law of 50 States to all innovations.”
Accordingly, “[t]he MDA expressly pre-empts only state requirements different from, or in addition to, any requirement applicable ... to the device under federal law.” Wolicki-Gables v. Arrow Int’l, Inc.,
“In order for a state requirement to be parallel to a federal requirement, and thus not expressly preempted under § 360k(a), the plaintiff must show that the requirements are ‘genuinely equivalent.’ State and federal requirements are not genuinely equivalent if a manufacturer could be held liable under the state law without having violated the federal law.”
B. Adequacy of Plaintiff’s Second Amended Complaint
Plaintiffs Second Amended Complaint asserts a single claim of negligence. See Doc. 44, p. 7. Plaintiff alleges that Defendant “fail[ed] to promptly and accurately report to the FDA incidents and problems with the EnPulse E1DR21 that occurred after the device received [PMA],” in violation of 21 U.S.C. § 360i, 21 C.F.R. § 803.50(a), 21 C.F.R. § 806.10(a), 21
Defendant argues that to the extent that Plaintiff alleges a breach of a duty owed to the Decedent, such a claim is expressly preempted by 21 U.S.C. § 360k(a). See Doc. 52, pp. 1-2. To the extent that Plaintiff alleges a breach of a duty owed to the FDA Defendant argues that such a claim is impliedly preempted by 21 U.S.C. § 337(a), as interpreted by the United States Supreme Court in Buckman Co. v. Plaintiffs’ Legal Comm.,
1. Express Preemption
In Riegel, the Supreme Court established a two-prong test for determining whether a state claim is preempted under the MDA: (1) determine whether the federal government has established requirements applicable to the medical device; and (2) if so, determine whether the state law claims are based upon requirements with respect to the device that are different from, or in addition to, the federal ones, and that relate to safety and effectiveness.
In her Response, Plaintiff asserts that “the essence of Plaintiffs claim is not that the defendant breached a duty to the FDA, but that it breached a duty to [the Decedent], who was in the foreseeable zone of risk created by Medtronic — namely, the duty to provide an adequate warning of the dangers of its product, which in this case paralleled its duty to comply with the conditions of its PMA.” Doc. 46, p. 11. For the same reasons discussed in the Court’s previous Order dismissing Plaintiffs Amended Complaint, such a claim would impose requirements that are different from, or in addition to, the federal requirements under the MDA. See Doc. 43, pp. 9-11. The MDA and its implementing regulations set forth a comprehensive scheme under which a device manufacturer must submit reports to the FDA regarding incidents where the device may have contributed to a death or serious injury. See 21 U.S.C. § 360i, 21 C.F.R. § 803.50; 21 C.F.R. § 814.84(a). Plaintiff has not pointed to any statute or regulation which imposes a duty on the manufacturer to inform patients about such incidents. The Court reiterates its position from its previous Order:
Under Plaintiffs theory of liability, Defendant would be required to provide information to ... patients beyond those on the device’s labeling. Accordingly, the state and federal requirements are not “genuinely equivalent” because Defendant could be held liable under the state law without having violated the federal law.
2. Implied Preemption
In addition to preempting non-parallel claims, the FDCA provides that all actions to enforce FDA requirements “shall be by and in the name of the United States.” 21 U.S.C. § 337(a). In Buckman, the United States Supreme Court construed § 337(a) as impliedly preempting suits by private litigants “for noncompliance with the medical device provisions.”
Riegel and Buckman create a narrow gap through which a plaintiffs state-law claim must fit if it is to escape express or implied preemption. The plaintiff must be suing for conduct that violates the FDCA (or else his claim is expressly preempted by § 360k(a)), but the plaintiff must not be suing because the conduct violates the FDCA (such a claim would be impliedly preempted under Buckman).
In re Medtronic, Inc., Sprint Fidelis Leads Prod. Liab. Litig.,
In Sprint Fidelis Leads, the plaintiffs alleged that the defendant had failed to file timely adverse event reports, as required by federal regulations.
Similarly here, to the extent that Plaintiffs negligence claim can be construed as alleging a breach of duty owed by Defendant to the FDA, such a claim is impliedly preempted by § 337(a). The thrust of Plaintiffs claim is that Defendant violated the FDCA and its implementing regulations by failing to inform the FDA about incidents where the E1DR21 may have contributed to a death or serious injury. See Doc. 44, ¶¶ 28-32. As with Plaintiffs claims in her Amended Complaint for negligence per se and failure to warn, Plaintiffs attempt to recast a claim for violation of the FDCA as a state-law negligence claim is impliedly barred by § 337(a). See Doc. 43, pp. 11-12; Sprint Fidelis Leads,
IV. CONCLUSION
After the Court dismissed Plaintiffs Amended Complaint, the Court granted Plaintiff leave to file a Second Amended Complaint, acknowledging that “pursuant to Riegel and Buckman, a narrow gap exists for a plaintiffs state law claim if it is to avoid express or implied preemption.” Doc. 43, p. 13. Having now again determined that Plaintiffs claim is expressly and impliedly preempted, the Court believes that any further opportunities to amend would prove futile. Accordingly, the Court will dismiss the Second Amended Complaint with prejudice.
Accordingly, it is hereby ORDERED and ADJUDGED:
1. Defendant Medtronic, Inc.’s Dispositive Motion to Dismiss Plaintiffs Second Amended Complaint (Doc. 45) is GRANTED.
2. Plaintiff Peggy McClelland’s Second Amended Complaint (Doc. 44) is DISMISSED with prejudice.
3. The Clerk is directed to terminate all pending motions and close this file.