MEMORANDUM OPINION AND ORDER
Plaintiff Christopher Hawkins filed this action claiming that he was injured by a medical device made by Defendant Medtronic, Inc. In his Amended Complaint, Plaintiff alleges that the device, a Medtronic Implantable Pulse Generator, Model # 7425 (hereinafter “IPG”), was defective and had to be replaced due to Defendant’s failure to satisfy its obligations under the Federal Food, Drug, and Cosmetic Act, as amended by the Medical Device Amendments of 1976, 21 U.S.C. § 301 et seq. (Am. Compl. ¶ 39.) Plaintiff filed this action asserting various causes of action under Ohio law. This matter is before the Court for consideration of Defendant’s Motion to Dismiss. (Dkt. 16.) For the reasons set forth below, the Court hereby GRANTS IN PART and DENIES IN PART Defendant’s Motion.
I. Background
Plaintiff asserts the following facts. On October 15, 2009, the IPG was surgically implanted into Plaintiffs back. The IPG is designed to send electrical pulses to the spinal cord in order to interfere with the transmission of pain signals and replace them with a tingling sensation called parasthesia. (Am. Compl. ¶ 9.) For a few months, the IPG worked as expected. (Id. at ¶ 12.) However, in late December 2009 or early January 2010, Plaintiff began experiencing painful shocks at the implant site, even when he occasionally turned off the device in an attempt to alleviate the shocks. (Id. at ¶¶ 13-14.) On numerous occasions, Plaintiffs’ doctors, with the assistance of a Medtronic agent, unsuccessfully attempted to fix the IPG. (Id. at ¶¶ 15-16.) On March 22, 2010, Plaintiffs doctor, after consulting with a Medtronic agent, determined that the IPG needed to be replaced. (Id. at ¶ 17.) On March 31, 2010, the IPG was excised and replaced by a new model. (Id. at ¶ 18.)
II. Preemption Under The Medical Device Amendments
Congress enacted the Medical Device Amendments of 1976 (“MDA”) “to provide for the safety and effectiveness of medical devices intended for human use.” Medtronic, Inc. v. Lohr,
At issue in this ease is the preemption provision of the MDA, contained in 21 U.S.C. § 360k(a), which provides generally
... no State ... may establish or continue in effect with respect to a device intended for human use any requirement—
(1) which is different from, or in addition to, any requirement applicable under this Act to the device, and
(2) which relates to the safety or effectiveness of the device or to any other matter included in a requirement applicable to the device under this Act.
The Supreme Court has held that this provision “simply was not intended to preempt most, let alone all, general common-law duties enforced by damages actions.” Medtronic, Inc. v. Lohr,
[sjtate requirements are pre-empted under the MDA only to the extent that they are “different from, or in addition to” the requirements imposed by federal law. Thus, § 360k does not prevent a State from providing a damages remedy for claims premised on a violation of FDA regulations; the state duties in such a case “parallel,” rather than add to, federal requirements.
Riegel v. Medtronic, Inc.,
To determine whether a state requirement is pre-empted, the Court first “must determine whether the Federal Government has established requirements applicable to” the device in question. Riegel,
Defendant correctly points out that “[cjlaims involving a PMA-approved device automatically satisfy [this] first condition of the preemption test.” (Dkt. 16-1 at 4.) In Riegel, the Supreme Court held that PMA “imposes ‘requirements’ under the MDA as [that Court] interpreted it in LoW’ because, “[u]nlike general labeling duties, [PMA] is specific to individual devices.” Riegel, 552 U.S. at 322-23,
If the federal requirements satisfy the first condition of the preemption test, the Court “must then determine whether [the plaintiffs] common-law claims are based upon [state] requirements with respect to the device that are ‘different from, or in addition to,’ the federal ones, and that relate to safety and effectiveness.” Riegel, 552 U.S. at 321-22,
Where the preemption issue is decided on the pleadings, and the complaint has not defined “the precise contours of [the plaintiffs] theory of recovery,” but “it is clear that the ... allegations may include claims that [the defendant] has, to the extent that they exist, violated FDA regulations,” the Supreme Court has held that these claims “can be maintained without being pre-empted by § 360k.” Lohr,
III. Standard of Review
Defendant seeks dismissal under Federal Rule of Civil Procedure 12(b)(6), which requires dismissal if the complaint fails to state a claim upon which relief can be granted. While Rule 8(a)(2) requires a pleading to contain a “short and plain statement of the claim showing that the pleader is entitled to relief,” in order “[t]o survive a motion to dismiss, a complaint must contain sufficient factual matter, accepted as true, to ‘state a claim to relief that is plausible on its face.’ ” Ashcroft v. Iqbal,
The context of MDA preemption affects the application of this standard, as explained by the Western District of Kentucky:
In the context of MDA preemption, Twombly and Iqbal make a plaintiffs job more difficult than it would be in a typical product liability case. When facing MDA preemption, a plausible cause of action requires, among other things, a showing that the alleged violation of state law parallels a violation of federal law This additional step requires some greater specificity in the pleadings. However, our appellate courts have beenunable to agree upon the precise level of that specificity. Nonetheless, ... a plaintiff must provide “more than labels and conclusions, and a formulaic recitation of the elements of a cause of action will not do.”
White v. Stryker Corp.,
IV. Plaintiffs’ Claims
Because Plaintiffs claims involve a PMA-approved device and therefore satisfy the first condition of preemption, the Court moves on to determine whether each claim against Defendant Medtronic is based on a parallel state requirement or one that is “different from, or in additional to” the federal requirements.
A. Count One: Design and Manufacturing Defects
In Count One, Plaintiff alleges that Defendant “designed, manufactured, distributed and sold the IPG [ ], an unreasonably dangerous, malfunctioning, and defective product to Ohio consumers, including Plaintiff Hawkins.” (Am. Compl. 41.) Plaintiff alleges that the IPG “contained a design and/or manufacturing defect, was adulterated, failed to comply with PreMarket Approval specifications, was not safe and/or effective, was non-conforming, and/or violated performance standards.” (Id. at 43.) Plaintiff alleges that this “defect rendered [the IPG] more dangerous than an ordinary consumer would expect when used in an intended or reasonably foreseeable manner.” (Id.) Plaintiff alleges that Defendant has therefore violated the statutes and regulations. listed below and is strictly hable for the resulting injuries (id. at 43, 44):
i. The following sections of the Ohio Revised Code:
a. Section 2307.74, which provides that “fa] • product is defective in manufacture or construction if, when it left the control of its manufacturer, it deviated in a material way from the design specifications, formula, or performance standards of the manufacturer, or from otherwise identical units manufactured to the same design specifications, formula, or performance standards.”
b. Section 2307.75, which provides generally that “a product is defective in design or formulation if, at the time it left the control of its manufacturer, the foreseeable risks associated with its design or formulation ... exceeded the benefits associated with that design or formulation.”
ii. The MDA, including but not limited to the following sections of Title 21 of the U.S.Code:
a. Section 351, which describes the circumstances under which a “device shall be deemed to be adulterated.”
b. Section 360(e), by which the Court assumes Plaintiff meant § 360e, which sets forth the. rules regarding the PMA process.
c. Section 360(d), by which the Court assumes Plaintiff meant § 360d, which sets forth the rules regarding performance standards.
d. Section 360(h), by which the Court assumes Plaintiff meant § 360h, which sets forth the rules regarding ■ the FDA’s authority to issue notifications regarding unsafe devices and provide other remedies.
iii. The following sections of Title 21 of the Code of Federal Regulations:
a. Section 820.1, which sets forth the scope of the FDA’s quality system regulation.
b. Section 820.5, which requires medical device manufacturers to establish and maintain a quality system.
c. Section 820.20, which provides for management responsibility.
d. Section 820.22, which requires quality audits.
e. Section 820.25, which sets qualification requirements for manufacturer personnel.
f. Section 820.30, which provides that manufacturers must “establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met.”
g. Section 820.70, which sets forth rules regarding production and process controls.
h. Section 820.72, which sets forth rules regarding inspection, measuring, and test equipment.
i. Section 820.75, which sets forth rules regarding process validation.
j. Section 820.80, which requires manufacturers to “establish and maintain procedures for acceptance activities ... including] inspections, tests, or other verification activities.
k. Section 820.86, which requires manufacturers to identify the “conformance or nonconformance of product with acceptance criteria ... throughout manufacturing, packaging, labeling, installation, and servicing of the product to ensure that only product which has passed the required acceptance activities is distributed, used, or installed.”
l. Section 820.90, which requires manufacturers to “establish and maintain procedures to control product that does not conform to specified requirements.”
m. Section 820.100, which requires manufacturers to “establish and maintain procedures for implementing corrective and preventive action” to address nonconforming product.
2. Design Defect
Defendant incorrectly argues that design defect claims are automatically preempted under the MDA. “[WJhere the FDA has specifically approved of the design of the device for investigational purposes, the Sixth Circuit has held that “[t]o allow a cause of action for design defect ... would thwart the goals of safety and innovation.” Martin v. Telectronics Pacing Sys.,
In Kemp, which was decided on summary judgment, “[t]he essence of plaintiffs’ [state law design defect claim was] that Medtronic failed to coat the lead” to a pacemaker to a uniform thickness. Kemp,
As in Kemp, the state law design requirements must not exceed the specific design requirements imposed through the PMA process. Because the preemption issue here must be decided on the pleadings, and the complaint has not defined “the precise contours of [the plaintiffs] theory of recovery,” the Court can not engage in a detailed comparison of the specific state and federal requirements at issue. However, because “it is clear that the ... allegations ... include claims that [Defendant] has ... violated FDA regulations,” these claims “can be maintained without being pre-empted by § 360k.” See Lohr,
2. Manufacturing Defect
Defendant asserts simply that “[a] manufacturing defect claim is undeniably preempted,” quoting Riegel for the proposition that “the MDA preempts] a negligent manufacturing claim insofar as it [is] not premised on the theory that [the defendant] violated federal law.” (Dkt. 16-1 at 7-8) (quoting Riegel,
While the complaint has not defined “the precise contours of [the plaintiffs] theory of recovery,” and the Court cannot engage in a detailed comparison of the specific state and federal requirements at issue, it is clear from the allegations that Plaintiffs claim is in fact premised on the theory that Defendant violated federal law. The Court therefore DENIES Defendant’s motion as to Plaintiffs manufacturing defect claim. See Riegel,
The Court notes that if, following the completion of discovery, it appears that Plaintiff cannot sustain a claim under state requirements that parallel federal requirements, Defendant would be free to file a motion for summary judgment.
B. Count Two: Failure to Warn
In Count Two, Plaintiff alleges that Defendant “knew or should have known of the potential for and/or actual presence of the defect,” and, “[h]aving such knowledge, Defendant failed to provide adequate warnings and/or instructions, both at the time of marketing and afterwards.” (Am. Compl. ¶ 49.) Plaintiff alleges that Defendant thus violated the following statutes and regulations (id. at 49-50):
i. Ohio Revised Code §§ 2307.76 and 2307.80, which provide a cause of action for product liability against a manufacturer which knows or should have known about a risk and fails to provide reasonable warning or instruction regarding the risk.
ii. 21 U.S.C. § 360(h), by which the Court assumes Plaintiff meant § 360h, which sets forth the rules regarding the FDA’s authority toissue1 notifications regarding unsafe devices and provide other remedies.
Hi. 21 U.S.C. § 360(i), by which the Court assumes Plaintiff meant § 360i, which requires manufacturers to establish and maintain records, make reports, and provide information to assure that devices are not adulterated or misbranded and to otherwise assure safety and effectiveness.
iv. 21 C.F.R. §§ 803.50 and 803.52, which describe reporting requirements.
v. 21 C.F.R. § 814.82, which describes allowable post-PMA requirements.
Defendant argues that Plaintiffs failure to warn claim is preempted, citing In re Medtronic Inc.,
To the extent that Plaintiffs allege that Defendants’ FDA-approved warnings “were inadequate under Ohio law,” such claims would be preempted. Kemp,
As with Count One, the complaint has not defined “the precise contours of [the plaintiffs] theory of recovery,” and the Court cannot engage in a detailed comparison of the specific state and federal requirements at issue. It is clear from the allegations, however, that Plaintiffs claim is premised on the theory that Defendant violated federal law. The Court therefore DENIES Defendant’s motion as to Count Two. See Riegel,
The Court notes that if, following the completion of discovery, it appears that Plaintiff cannot sustain a claim under state requirements that parallel federal requirements, Defendant would be free to file a motion for summary judgment.
C. Count Three: Negligent Handling
In Count Three, Plaintiff alleges that Defendant “failed to establish and/or maintain adequate distribution, installation, handling, and/or inspection instructions and/or procedures” in violation of the MDA and 21 C.F.R. §§ 820.140, 820.150, 820.160, and 820.170. (Am. Compl. ¶ 54.)
Because Plaintiff does not identify a cause of action under Ohio law, and the MDA provides none, Lohr,
D. Count Four: Breach of Express and Implied Warranties
In Count Four, Plaintiff alleges that “Defendant breached its expressed and implied warranties regarding the safety and utility of its defective IPG [ ].” (Am. Compl. ¶ 58.)
1. Express Warranty
Defendant argues that Plaintiffs claim of breach of express warranty is preempted, citing the Fifth Circuit’s decision in Gomez v. St. Jude Med. Daig Div., Inc.,
Here, however, a claim for breach of express warranty does not require a finding that the representations are untrue. The elements of such a claim under Ohio law are as follows:
(1) the manufacturer of the product, through advertising, makes representations regarding the quality and merit of its product,
(2) the representations are aimed directly at the ultimate consumer, urging the consumer to purchase the product,
(3) the consumer, relying on the manufacturer’s representations, does purchase the product, and
(4) the consumer suffers harm as a result of that reliance.
Wagner v. Roche Lab., 85 Ohio St.3d 457,
Defendant develops no other arguments as to this claim, and its motion is DENIED as to Plaintiffs claim for breach of express warranty.
2. Implied Warranty
The elements of an Ohio claim for breach of implied warranty claim are:
(1) the existence of a defect in the product manufactured and sold by the defendant,
(2) the defect existed when the product left the hands of the defendant, and
(3) the defect was the direct and proximate cause of the plaintiffs injuries.
Tompkin v. Philip Morris USA, Inc.,
According to Defendant, the Riegel Court held that “the MDA pre-empt[s] claims of ... breach of implied warranty.” (Dkt. 16-1 at 10 (quoting Riegel,
The Court notes that if, following the completion of discovery, it appears that Plaintiff cannot sustain a claim under state requirements that parallel federal requirements, Defendant would be free to file a motion for summary judgment.
E. Count Five: Negligent and Fraudulent Misrepresentation
In Count Five, Plaintiff alleges that “Defendant provided a defective product which failed to conform to [Defendant’s] representations,” and that Defendant “negligently and/or fraudulently misrepresented the safety and utility of’ the IPG, in violation of Ohio Revised Code § 2307.77 and 21 C.F.R. § 801.6. (Am. Compl. ¶ 63.)
The Sixth Circuit has held that, in the context of investigational devices, claims under Ohio Revised Code § 2307.77 are preempted by the MDA. Martin v. Telectronics Pacing Sys.,
F. Count Six: Failure to Report
In Count Six, Plaintiff alleges that Defendant “fail[ed] to adequately and accurately record and/or report a malfunction, adverse event, reportable event, and/or product problem associated with [the IPG],” in violation of 21 U.S.C. § 360(i) and 21 C.F.R. §§ 803.1, 803.10, 803.50, 803.52, 814.82, 820.65, 820.90, 820.184, 820.186, 820.198, 820.200, and 821.25. (Am. Compl. ¶ 68.) Plaintiff alleges that Defendant “failed to have a system to identify defective products, to maintain a quality system record, [and] to maintain a device history reeord[;] and failed to properly investigate and/or report a complaint or service a device.” (Id.)
The MDA provides no private cause of action, Lohr,
Y. Conclusion
For the reasons discussed above, Defendant’s Motion is GRANTED as to Counts Three, Five, and Six and DENIED as to
IT IS SO ORDERED.
Notes
. Section 360k(b) provides that exemptions may be granted to states under certain circumstances.
. In Lohr, Medtronic "argue[d] that the plain language of the statute pre-empts any and all common-law claims brought by an injured plaintiff against a manufacturer of medical devices.” Lohr,