Blinn v. Smith & Nephew Richards, Inc.Blinn v. Smith & Nephew Richards, Inc.
ORDER ON DEFENDANT’S MOTIONS FOR SUMMARY JUDGMENT
This cause comes before the Court on the following:
1. Defendant SMITH & NEPHEW RICHARDS, INC.’s motion for partial summary judgment on the issue of IDE preemption and supporting memorandum (Docket Nos. 23-24), and Plaintiff MARIS BLINN’s response (Docket No. 28); and
2. Defendant’s motion and memorandum for partial summary judgment on manufacturing defect and regulatory violations (Docket No. 36), and Plaintiffs response (Docket No. 46).
STANDARD OF REVIEW
Summary judgment is proper if “the pleadings, depositions, answers to interrogatories, and admissions on file, together with the affidavits, if any, show that there is no genuine issue as to any material fact and that the moving party is entitled to a judgment as a matter of law.” Fed. R.Civ.P. 56(c). The moving party bears the burden of proving that no genuine issue of material fact exists.
See Celotex Corp. v. Catrett,
The evidence presented must be construed in favor of the non-moving party, and that party must receive the benefit of all favorable inferences that can be drawn from that party’s evidence.
United States v. Diebold, Inc.,
BACKGROUND
I. Statutory Framework
The Medical Device Amendments [MDA] of the Food, Drug, and Cosmetic Act [FDCA], 21 U.S.C. Section 360, make the Food and Drug Administration [FDA] responsible, with the aid of non-governmental experts, for classifying devices entering the market into categories in order to determine the level of regulation necessary to ensure their safety and effectiveness. The MDA divide medical devices into three categories. Class I devices are subject only to the “general controls” that apply to medical devices in all three classes. 21 U.S.C. § 360c(a)(l)(A). Class I devices include items such as tongue depressors, elastic bandages, ice bags and bed pans. See Robert Adler, The 1976 Medical Device - Amendments: A Step in the Right Direction Needs Another Step in the Right Direction, 43 Food & Drug L.J. 511, 513 (1988). Class II devices are subject to “special controls,” such as performance standards, patient registries, and the dissemination of guidelines. 21 U.S.C. § 360c(a)(l)(B). Class II devices include bone plates, hearing aids, and syringes. See Adler, 43 Food & Drug L.J. at 513. Class III devices are the most heavily regulated. Some examples of Class III devices are pacemakers, intrauterine devices, artificial hearts and artificial joints. See Adler, 43 Food & Drug L.J. at 513.
There are two paths by which Class III devices may generally be approved for marketing. The first, pre-market approval [PMA], involves more extensive regulation. It requires the manufacturer to provide a “reasonable assurance” of the product’s safety and effectiveness. 21 U.S.C. § 360c(a)(1)(C). This is a “rigorous” process, in which the “[m]anufacturers must submit detailed information regarding the safety and efficacy of their devices, which the FDA reviews, spending an average of 1,200 hours on each submission.”
Medtronic, Inc. v. Lohr,
Devices that are not approved for general marketing may still reach the market on a limited basis through the investigational device exemption [IDE].
See
21 U.S.C. 360j(g). The objective of the IDE is “to encourage, to the extent consistent with the protection of the public health and safety and with ethical standards, the discovery and development of useful devices intended for human uses and to that end to maintain optimum freedom for scientific investigators in their pursuit of that purpose.” 21 CFR § 812.1(a). “Although
the risks to the subjects are not outweighed by the anticipated benefits to the subjects and the importance of the knowledge to be gained, or informed consent is inadequate, or the scientific investigation is scientifically unsound, or there is reasons to believe that the device as used is ineffective.
21 C.F.R. § 812.30. IDEs are specifically exempted from many of the requirements to which they would otherwise be subject as Class III medical devices. See 21 C.F.R. § 812.1(a).
II. Facts
This case is one of approximately two thousand products liability actions against this Defendant and other manufacturers of orthopedic bone screws. These actions arise out of the surgical implantation of these devices in the pedicles of the spine. The bone screw devices are used in conjunction with spinal fusion surgery on patients with a number of severe back problems. They are intended to stabilize the spine to allow spinal fusion to take place.
The case was transferred to the United States District Court for the Eastern District of Pennsylvania by the Judicial Panel on Multidistrict Litigation and placed on the docket of MDL 1014, In re Orthopedic Bone Screw Products Liability Litigation, before the Honorable Louis C. Bechtle. Following pretrial proceedings and discovery, the case was remanded to this Court for disposition.
The facts as stated are taken from Plaintiffs complaint and the exhibits to Plaintiffs and Defendant’s pleadings. Defendant manufactures the Rogozinski Spinal Rod System, a bone screw device. In 1990, Defendant applied for an IDE for use of the Rogozinski System in the pedi-cles of the spine. At that time, Defendant already had § 510(k) clearance for use of the Rogozinski System in the lumbar spine and the sacrum. The FDA required a number of changes and corrections in Defendant’s application, which Defendant made. The FDA gave final approval to Defendant’s IDE application in August 1991.
Plaintiff suffers from degenerative disc disease. She was identified by Dr. Rogo-zinski as a potential participant in the IDE study. On March 19,1992, Plaintiff signed the consent form the FDA authorized for use in the Rogzoinski System IDE study. The form states that the device is experimental. It identifies the following risks of use of the Rogozinski System:
a. All surgical procedures carry the possibility of blood loss, anesthetic complications, deep vein thrombosis (blood clots), blood transfusion reactions, nerve injury, and pneumonia. Some of the possible complications from spinal surgery are: fracture of the bony structure of the spine, failure and/or loosening of the instrumentation, infection, nerve root and/or spinal cord injury, allergic reaction, arterial compromise, and metal ion sensitization or toxicity.
b. The possible complications are not limited to the above and it is not possible to advise one of every conceivable complication. Those outlined are cited to be reasonably known risks of spinal surgery and are presented so your decision may be made on a knowledgeable foundation.
c. The following are risks to which you may be exposed during this investigation: failure of spinal fusion to occur, failure and/or loosening of the instrumentation even in the presence of fusion.
The form additionally outlines the discomforts associated with the operation and the alternatives available, and states that participation in the study is voluntary.
Plaintiff underwent spinal surgery on March 26, 1992, during which the Rogozin-ski System was implanted. Spinal fusion, the intended result of the surgery, occurred. However, complications from the original surgery developed, and Plaintiff had to have a number of further surgeries. Additionally, the site of the surgery became infected, and has since developed a chronic infection.
ANALYSIS
Plaintiffs complaint consists of a single count, labeled “Claim Against Smith & Nephew Richards” that appears to include claims on multiple theories, without labeling any of those claims. From Plaintiffs complaint and pleadings it appears that Plaintiff intended to plead breach of warranty, failure to warn, manufacturing defect, and negligence per se claims. The Court will therefore assume that these are the theories under which Plaintiff proceeds. Defendant argues that summary judgment is appropriate with respect to the failure to warn and the breach of warranty claims on the ground that they are preempted by the IDE. With respect to the claims based on manufacturing defect and regulatory violations, Defendant moves for summary judgment on the ground that there are no disputed issues of material fact and Defendant is entitled to judgment as a matter of law.
I. IDE Preemption
Defendant argues that claims for breach of warranty and failure to warn on a device subject to an IDE are pre-empted. “By virtue of the Constitution’s Supremacy Clause, it long has been settled that ‘state law that conflicts with federal law is without effect.’ ”
Goodlin v. Medtronic, Inc.,
establish or continue in effect with respect to a device intended for human use any requirement—
(1) which is different from, or in addition to, any requirement applicable under this chapter to the device, and
(2) which relates to the safety or effectiveness of the device or to any other matter included in a requirement applicable to the device under this chapter.
21 U.S.C. § 360k(a). Although neither case decided the precise issue before this Court, the reasoning of the Supreme Court’s decision in
Lohr
and the Eleventh Circuit’s decision in
Goodlin v. Medtronic, Inc.,
A. The Lohr decision
In Lohr, the Supreme Court considered the extent to which state law claims regarding a device approved by the FDA under the pre-market notification process are preempted by federal law. The Court stated that its analysis of that question:
is guided by our oft-repeated comment, ... that the purpose of Congress is the ultimate touchstone in every pre-emption case. As a result, any understanding of the scope of a pre-emption statute must rest primarily on a fair understanding of congressional purpose. Congress’ intent, of course, primarily is discerned from the language of the preemption statute and the “statutoryframework” surrounding it. Also relevant, however, is the structure and purpose of the statute as a whole, as revealed not only in the text, but through the reviewing court’s reasoned understanding of the way in which Congress intended the statute and its surrounding regulatory scheme to affect business, consumers, and the law.
The Lohr Court then considered whether the negligent design claims asserted by the plaintiff in that case were preempted. It noted that the approval process for a device proceeding under the pre-market notification provisions “is focused on equivalence, not safety.” Id. at 2254. Thus, although the FDA might have examined the § 501(k) application “with a concern for the safety and effectiveness of the device,” the FDA did not require to device “to take any particular form for any particular reason.” Id. The Court also examined the Congressional purpose behind the § 501(k) process. That purpose was “merely to give manufacturers the freedom to compete, to a limited degree, with and on the same terms as manufacturers of medical devices that existed prior to 1976.” Id. The Court therefore held that state law negligent design claims with respect to a device approved under the § 501(k) process do not impose any additional requirements different from or in addition to those imposed by the MDA, and are not preempted. Id. Using the same analysis, the Court concluded that a claim based on violation of FDA regulations would not be pre-empted. See id. at 2256.
The Court then addressed the question of whether a failure to warn claim was preempted. The federal regulation in question “requires manufacturers of every medical device, ... to include with the device a label containing ‘information for use, ... and any relevant hazards, contraindications, side effects, and precautions.’ ” Id. at 2256 (quoting 21 C.F.R. §§ 801.109(b) and (c)). The Court stated that “the federal requirements reflect important but entirely generic concerns about device regulation generally, not the sort of concerns regarding a specific device or, field of device regulation which the statute or regulations were designed to protect from potentially contradictory state requirements.” Id. at 2258. It noted as well that “the general state common law requirements in this case were not specifically developed ‘with respect to’ medical devices.” Id. The Court stated that:
the predicate for the failure to warn claim is the general duty to inform users and purchasers of potentially dangerous items of the risks involved in their use. These general obligations are no more a threat to federal requirements than would be a state-law duty to comply with local fire prevention regulations and zoning codes, or to use due care in the training and supervision of a workforce. These state requirements therefore escape pre-emption ... because their generality leaves them outside the category of requirements that § 360k envisioned to be “with respect to” specific devices such as pacemakers.
Id. at 2258.
B. The Goodlin decision
In
Goodlin,
the Eleventh Circuit addressed the extent to which state law claims are preempted with respect to a medical device approved under the more rigorous PMA process. The Eleventh Circuit interpreted
Lohr
as meaning that a state law claim would be preempted by section 360k(a) if three requirements were met: “(1) the imposition of a specific federal requirement that (2) applied to a particular device and (3) focused on the safety and effectiveness of the device.”
Goodlin,
The approval represents only a finding that the manufacturer’s proposal to market a device has reasonably assured the FDA of the device’s safety and effectiveness. Nor does the FDA’s willingness to notify an applicant of deficiencies and to propose modifications to the PMA application add any further force to [the] argument for preemption, because an applicant who corrects or modifies a deficient PMA application before receiving the FDA’s approval stands ip no better position that an applicant whose initial PMA application was flawless. In either case, the FDA enters a finding that the applicant has furnished the relevant assurances and therefore may begin to market its device. In neither case, however, does the approval provide any indication of what (if any) specific substantive requirements the FDA may have applied to reach that result.
Id.
The Eleventh Circuit recognized that its decision regarding the preemptive effect of PMA approval was at odds with the decisions of other courts. Nonetheless, it stated that its holding was supported by the language and structure of the MDA. The stated purpose of the MDA is “provid[ing] for the safety and effectiveness of medical devices intended for human use.” Id. at 1378 (quoting Pub.L. No. 94-295, 90 Stat 539, 539 (preamble)) (alteration in original). Moreover, passage of the MDA was prompted by “several highly publicized incidents involving defective medical devices.” Id. “It would have been inconsistent for the same Congress that enacted these sweeping reforms to make a potentially dangerous industry safer for patients by blocking the admission of defective devices to the market, then to preempt product liability suits when those devices caused injury.” Id.
C. Failure to Warn
Neither the Supreme Court nor the Eleventh Circuit has addressed the issue of preemption with regard to devices introduced under an IDE. The federal courts that have considered this issue post-Lofcr have uniformly held that there is at least some preemptive effect to an IDE.
See Chambers v. Osteonics Corp.,
Plaintiff alleges that she was not given sufficient warning regarding the potential for infection caused by implantation of the Rogozinski System. In
Lohr,
the Supreme Court examined such a claim, and stated that a claim based on “the general duty to inform users and purchasers of potentially dangerous items of the risks involved in their use” was not preempted
Defendant argues that failure to warn claims are nonetheless preempted in the IDE context for two reasons. First, Defendant argues that the modifications the FDA required on the label of the Rogozin-ski System constitute device-specific requirements. As discussed above, in the time since Defendant filed its brief, the Eleventh Circuit rejected this argument in the context of devices subject to the PMA process.
See Goodlin,
Second, Defendant argues that the FDA regulations regarding IDEs evidence an intent to exempt IDEs from claims such as Plaintiffs. The IDE regulations expressly exempt IDEs from the labeling requirements the Supreme Court discussed in the context of failure to warn claims. See 21 C.F.R. § 812.1(a). The analytical framework of Defendant’s argument is thus somewhat different than the analysis used in Lohr or in Goodlin, because the argument is not based on the presence of a particular federal regulation with which a state law claim is alleged to conflict. Instead, Defendant’s argument is built on the fact that the IDE regulations provide for exemption from particular federal regulations. However, this Court finds Defendant’s argument convincing.
As the Supreme Court stated in
Lohr,
“the purpose of Congress is the ultimate touchstone in every pre-emption case.”
Permitting a failure to warn claim to stand when there is uncontroverted evidence that the manufacturer complied with the IDE regulations regarding consent and labeling would “permit a factfinder to adjudge as inadequate the FDA approved labeling and warning language, the FDA approved consent form, and the investigator agreements signed by approved FDA investigators.”
Touchet,
D. Breach of Warranty
Although Defendant moved for summary judgment as to the breach of warranty count on the ground of preemption, this Court does not find it necessary to address the question of whether such a claim is preempted. Plaintiffs claim for breach of an express warranty was dismissed by Judge Bechtle. In Pretrial Order Number 651, the MDL court ordered that all claims for breach of express warranty be dismissed with prejudice, except where plaintiffs responded to a show cause order requiring plaintiffs to identify the specific express warranty at issue, describing how it was relied on, and specifying the manner in which it was allegedly breached. Plaintiff made no such response. To the extent Plaintiffs breach of warranty claim
II. Absence of Genuine Issues of Material Fact
A. Product Defect
Plaintiff presents no evidence supporting a claim based on product defect, under either a negligence or a strict liability theory. Under either theory, Plaintiff would be required to provide expert testimony that the product was defective and evidence that the product caused the injury of which she complains.
See Alexander v. Danek Medical, Inc.,
B. Negligence Per Se
Defendant argues that Plaintiff fails to establish a negligence per se claim based on any regulatory violation because Plaintiff fails to identify any applicable statute or regulation Defendant violated, and because Florida law does not allow the use of a statutory violation as the basis for civil liability unless the statute expressly provides a private cause of action. In its response, Plaintiff again fails to specifically identify a statute or regulation Defendant allegedly violated. Plaintiff does refer in general terms to violations of the FDA’s regulations regarding informed consent and regarding marketing and promotion, presumably referring to the regulations promulgated by the FDA under the authority granted to it by the FDCA.
Assuming that the alleged regulatory violation under which Plaintiff proceeds is a violation of FDA regulations, a negligence per se claim is not available to Plaintiff. Under Florida law, whether there is civil liability for violation of a statute depends on legislative intent.
See Murthy v. N. Sinha Corp.,
In any event, even under a negligence
per se
theory, a plaintiff must prove causation.
See Groh v. Hasencamp,
ORDERED that Defendant SMITH & NEPHEW RICHARDS, INC.’s motions
DpNE AND ORDERED.